Tracking laser tomography

This article is best viewed in a PDF Format.

View PDF 

 Get adobe

Optical coherence tomography (OCT) is a non-contact and non-invasive high resolution technique for imaging of partially transparent objects. It has found a wide spectrum of applications in biomedical imaging, especially in ophthalmology.1,2

The technique is analogous to ultrasound, except that it uses light waves rather than sound waves. OCT offers information about the depth structure of a sample using interferometry of temporally low coherent light at a wavelength of approximately 840nm. Since the introduction of the OCT in 1991, this revolutionary technology has constantly improved its design with more advanced imaging modalities and acquisitions. There are two variants of OCT techniques depending on the detection system: time-domain, such as the Stratus OCT 3 (Carl Zeiss Meditec) and frequency-domain as found in the Spectralis HRA+OCT (Heidelberg Engineering, Haag-Streit UK distribution). Frequency-domain provides a significant improvement of imaging speed and detection sensitivity.3 The aim of this paper is to show the advanced applications of the Spectralis HRA+OCT in a clinical setting.

The Spectralis

The Spectralis HRA+OCT (Figure 1) provides a unique combination of retinal angiography and optical coherence tomography. The Spectralis HRA+OCT comes with six different acquisition modes (Figure 2). OCT, fluorescein angiography (FA) and indocyanine green angiography (ICGA) are the three main ones. Additionally, it enables multimodality imaging where any two modalities can be used simultaneously. Simultaneous dual-beam imaging also allows infrared fundus imaging, where high resolution black and white images reveal advanced details and structures of the inner eye. The red-free and fundus-autofluorescence modalities are especially designed to enhance imaging during fluorescein angiography and indocyanine green angiography. For optometrists in a clinical setting, the retinal angiography modes might not be a necessity Heidelberg also offers Spectralis versions without this mode (Spectralis OCT or Spectralis OCTplus). New software has been released with features listed in Table 1.

Eye tracking

The Spectralis HRA+OCT has one main advantage compared to other available OCT devices. It uses an integrated confocal laser fundus imaging technique to guide the OCT scan and introduce eye tracking, by using two separate beams of light (820 and 870µm) to capture two images simultaneously at 40,000 A-scans per second. One beam acts as a reference, guiding the second beam of light during a real time, cross-sectional scan and keeping it correctly positioned as the eye moves. This technique enables high precision and repeatability. Compared to other available OCT instruments, the Spectralis HRA+OCT presented very low coefficients of variation for repeated measurements, which makes the instrument highly repeatable. Additionally, it showed the smallest variability, which equates to the smallest measurable change of 3µm, when compared to other OCT instruments (between 7 and 16µm).4

Axial resolution has shown the greatest improvement over time with various imaging techniques. The axial resolution of the Spectralis instruments achieve a 7.0µm optical resolution and a 3.9µm digital resolution, which is again an improvement compared to the 3D OCT-1000 of Topcon (5µm). Figure 3 and Figure 4 show cross-sectional images obtained with the Spectralis, showing all retinal layers in high definition. The Spectralis outperforms previously developed 3D OCT instruments in terms of the scanning speed (0.019 compared with 0.05 seconds per scan).

Other additional advantages to this new technique include tracking changeover time and noise reduction. The instrument will not only use the baseline fundus image as its guide during the first OCT scan, but uses its location in any follow-up scans as well. It will automatically place any future scans in the exact same location, which eliminates any subjective placement error (Figure 5). This feature will improve its disease assessment over time and helps to improve any treatment decisions. The noise reduction, also known as the real-time (ART) function, produces higher quality images with finer details, based on combining multiple images of the same location and filters out any random noise which was not present in every image in the entire dataset. The confocal microscope, which suppresses the signal anteriorly and posteriorly to the confocal plane of interest, will also reduce the interference with anterior anatomical structures such as the crystalline lens.

Use in practice

The use of the Spectralis instrument is fast and easy for both the observer and the patient. Due to the eye tracking system, the acquisition of the OCT scan might take a bit longer than other frequency domain OCTs. No dilation is needed while the patient is asked to place their chin and head against the rest and look at the blue internal fixation target. The following parameters can be changed on the control panel: intensity of IR (25-100 per cent in steps of 25 per cent) width of the scan (15, 20, or 30 degrees) and either a section scan (single line, circular, radial) or a volume scan. The Spectralis allows the operator to customise the volume scanning protocol by adjusting the scan density and the dimensions of the scan area. From each volume scan, the Spectralis generates a default retinal thickness map consisting of three concentric circles measuring 1, 2, and 3mm in diameter but allows the user to choose from several map displays, which automatically recalculate retinal thickness measurements. There is an option to adjust the eye length for patients with myopic or deep set eyes. However, patients with >12D prescription should be asked to wear their glasses.

Output data

The output of the OCT scan (Figure 6) looks very similar to other OCT instruments and uses the same colour coding. Three-dimensional volume scans are becoming the standard imaging method for assessing the retina. The Spectralis instruments are pre-installed with remote viewing capabilities that enable clinicians to review images at their PC or even display them on large monitors for educating patients. The software produces a retinal thickness map, RNFL thickness map, and progression map. These images can be displayed for example in colour spectrum red (enhanced viewing of intra-retinal layers) or blue (enhanced detail around the RPE). Other viewing options are available, such as colour (heat and high frequency) or grey scales (positive and negative), which can all be viewed in 2D or 3D.

Conclusion

Management and treatment of retinal diseases has improved dramatically over the past several decades and this has mainly occurred as imaging techniques became more advanced. Retinal thickness changes and the presence of fluid are the two most important parameters in the follow up of retinal diseases, which needs high reproducibility of any imaging technique. The Spectralis HRA+OCT has shown high repeatability and reproducibility in normal subjects when compared to other available OCT devices, with dramatic advances in virtual every aspect of imaging retinal diseases, due to its multimodality imaging, eye-tracking and noise reduction. ●

Acknowledgement

Thanks to Haag Streit UK for loan of the instrument to the Fight for Sight Optometry Clinic, City University. For more details on Spectralis telephone Haag Streit on 01279 456261.

References

  1. Schmidt-Erfurth U, Leitgeb RA, Michels S, et al. Three-dimensional ultrahigh-resolution optical coherence tomography of macular diseases. Invest Ophthalmol Vis Sci, 200546:3393-3402.
  2. Christopoulos V, Kagemann L, Wollstein G, et al. In vivo corneal high-speed, ultra high-resolution optical coherence tomography. Arch Ophthalmol, 2007125:1027-1035.
  3. de Boer JF, Cense B, Park BH, Pierce MC, Tearney GJ, Bouma BE. Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography. Opt Lett, 200328:2067-2069.
  4. Wolf-Schnurrbusch UE, Ceklic L, Brinkmann CK, et al. Macular thickness measurements in healthy eyes using six different optical coherence tomography instruments. Invest Ophthalmol & Vis Sci, 200950:3432-3437.

● Dr Byki Huntjens is a lecturer at City University and has no commercial interests in any product described




Spread the word:   bookmark it! diggit! reddit!




Optician magazineProviding exclusive eye care news, information and educational needs every week, including a FREE CET programme. Subscribe to Optician Print Edition.

Who won?

Those Optician Award winners in full

OFFERS OF THE WEEK

Click here to find out more about the products

Email newsletter

Sign up for our fortnightly email newsletter by clicking here.

BUY OPTICIAN

Get a copy of Optician delivered to your door every week.